MPA1 Instruction Manual Page 30
Calibration
CAUTION: ROUTINE CALIBRATION
SHOULD NOT BE REQUIRED. THE
FOLLOWING PROCEDURE SHOULD ONLY
BE USED BY A QUALIFIED TECHNICIAN.
This procedure will restore the MPA1 to
factory calibration if controls have been
moved from factory settings.
+48V Adjustment
1. Using a DC voltmeter, measure the
voltage across D23 and adjust “+48V
Adjust” (R107) until the meter reads
48V.